Make: Smart Instruments
Model:SMART SORB 93
Specification /Features:
- Surface Area Range – 0.1m2/gm to 1500 m2/gm Normal
- Accuracy – Typically better than ±5%
- Reproducibility- Typically better than ±3% Multi Point (5 points) Surface Area can
be measured with additional mixture gas cylinder
- Total Pore Volume can be measure with separate mixture gas
- Sample Holder – Typical Sample Holder Capacity 7ml
- Regeneration System -: Temperature Range
- Ambient to 3000C Accuracy: Better than ±50C of Set Point
TOTAL COST: 9 LAKHS
Applications
The SMART SORB 93 BET Surface Area Analyzer measures the specific surface area, pore size,
and pore volume of materials, crucial for characterizing adsorbents, catalysts, and
nanomaterials. It has a surface area range of 0.1 m²/g to 1500 m²/g with accuracy better
than ±5% and reproducibility better than ±3%. The instrument supports multi-point surface
area measurements and total pore volume analysis using additional gas mixtures. Its sample
holder accommodates 7 ml, and the built-in regeneration system operates from ambient to
300°C with high precision. This analyzer is widely used in materials science,
pharmaceuticals, and environmental studies to optimize material performance and ensure
quality control.
FOURIER TRANSFORM-INFRARED
SPECTROPHOTOMETER (FT-IR) WITH ATR ZnSe
Make: Bruker, Germany
Model: Alpha-II
Specification /Features: -
- Spectral range should be from 6000cm-1 to 500 cm-1 which covers metal
oxides in Mid
IR Range.
- Interferometer should be permanently aligned (No mechanical or dynamic
alignment is
preferred.) with 10 Years Warranty
- For High Humidity Beam Splitter and Windows: ZnSe
- Universal ATR for direct sampling for Solid, Liquid, Gum, Paste and Film
is
required.
- MIR Source with 5 Years warranty
- Laser Source with 10 Years Warranty
- Spectral resolution should be 2cm-1.
- System should have Gold coated mirrors.
- Wave Number Accuracy: 55000:1
- Photometric accuracy: better than 0.1%T
- Signal to Noise ratio: >55000:1
- Detector: High sensitivity DLATGS room temperature with analog to
digital conversion
with 24 Bit energy for stability.
- Computer interface: Ethernet, remote control via W-LAN (Integrated PC
remote
control if possible).
TOTAL COST 15 LAKHS
Applications
The Fourier Transform Infrared (FT-IR) Spectrophotometer Alpha-II by Bruker,
Germany, is
designed for advanced material analysis with features ideal for mid-IR
applications. It
operates across a spectral range of 6000 cm⁻¹ to 500 cm⁻¹, making it
suitable for
studying metal oxides and other materials. The system includes a permanently
aligned
interferometer with a 10-year warranty and uses ZnSe components for high
humidity
resistance. It offers universal ATR sampling for solids, liquids, pastes,
gums, and
films with a spectral resolution of 2 cm⁻¹ and photometric accuracy better
than 0.1%T.
With a high-sensitivity DLATGS detector, signal-to-noise ratio >55000:1, and
gold-coated
mirrors, it ensures precision and stability. Additionally, it features
Ethernet and WLAN
capabilities for remote control.
HIGH-PERFORMANCE LIQUID CHROMATOGRAPHY(HPLC) WITH PDA
DETECTOR
Make: Waters USA
Model: 515 PUMP with EM Power QA software
Specification /Features: -
- System Capability: The System must be capable of to carry out
analytical
applications
& upgradable to semi preparative analysis in future.
- Pump: High Pressure Dual Pump Binary Gradient System with
analytical capability
Programmable flow range: 0.001 – 10 ml/min Delay volume with mixer should be
≤200 or
equivalent Flow accuracy: ± 0.1% (for individual pump) or better Flow precision:
< 0.1% RSD (for individual pump) Operating pressure: 5000 psi or more Pressure
ripple of single pump-≤2% at 1.0 ml/min
- Injector: Manual, Analytical injector system
-
Photo Diode Array Detector: PDA detector with Analytical flow cell
desired for
detection of UV-Chromophoric compounds. Its uniquely integrated software and
optics
innovation deliver no compromise chromatographic and spectral sensitivity, and
has a
noise specification as low as 10uAU. Utilized for compound confirmation method
development, or routine analysis, the PDA is reliable and easy-to-use. Software
allow flexibility of either 2D or 3D operation for spectral comparison and peak
tracking.
Parameter
|
Specification
|
Light source
|
D2 lamp or
better
|
Wavelength range
|
200-800 nm or
better
|
Flow cell
Volume
|
8.4 μL or
better
|
Lamp Replacement
|
Front access or
better
|
Lamp Life Time
|
2000hours (D2 lamp) or
better
|
5. Analytical Column for Method Development:
C18 dimension 4.6mmx 250mm (5 Micron) – Quantity 1
6. Software: Windows based, with a relational secured data
base, independent of
operating
system like SQL/Oracle, with strong integrity and security of data. All meta data should
be
automatically managed, linked and versioned by the software. Software should be able to
programme multiple gradient curves simultaneously, with apex integration and Gaussian
skimming. The software should be compatible with the PC and the instrument.
TOTAL COST: 25 LAKH
Applications
The High-Performance Liquid Chromatography (HPLC) system with PDA Detector by Waters USA,
Model
515 Pump, is designed for high-precision analytical applications, with upgradability to
semi-preparative analysis in the future. It features a high-pressure dual pump binary
gradient
system with a programmable flow range of 0.001 – 10 ml/min, flow accuracy of ±0.1%, and a
maximum operating pressure of 5000 psi or more. The system ensures exceptional performance
with
a manual analytical injector and a Photo Diode Array (PDA) detector offering a wavelength
range
of 200–800 nm, noise specification as low as 10 μAU, and robust spectral sensitivity for
method
development and routine analysis.
The analytical column (C18, 4.6mm x 250mm, 5 μm) supports method development, while the EM
Power
QA software provides secure, Windows-based operation with SQL/Oracle compatibility, ensuring
robust data integrity and versioning. The software allows flexibility for 2D/3D spectral
comparison, peak tracking, and advanced gradient programming. With features like a D2 lamp
with
a lifetime of 2000+ hours and an 8.4 μL flow cell, this HPLC system is reliable, efficient,
and
easy to use, ensuring precise chromatographic results for diverse applications.
UV/VIS SPECTROPHOTOMETER
Make: LABINDIA
Model: UV 3092
Specification /Features:
- True Double Beam with Photo Multiplier Tube Detector System.
- Detector: Photo Multiplier Tube Detector
- Wavelength range: 190-900nm.
- Lamps/Source Required: Tungsten and Deuterium
- Built-in Automatic dual cell changer
- Stray light less than 0.01%.
- Photometric Range - -4 to +4.
- Wavelength Accuracy: +/- 0.1nm at 656.1nm
- Baseline Flatness: ±0.0006 Abs or better
- Scan Speed – Slow/Medium/Fast Selectable
- Variable Bandwidth from 0.1nm to 5 nm Continuous.
- Instrument should have facility to work with SOILD SAMPLES like metal oxides and
metal powder with Integrating Sphere.
TOTAL COST : 6.5 LAKHS
Applications
The UV/VIS Spectrophotometer by LABINDIA, Model UV 3092, is a high-precision analytical
instrument designed for advanced spectroscopic applications. It features a true
double-beam
system with a Photo Multiplier Tube Detector for exceptional sensitivity and accuracy.
The
instrument offers a wavelength range of 190–900 nm, covering applications in UV and
visible
spectroscopy, and uses Tungsten and Deuterium lamps for reliable light sources.
Key features include a built-in automatic dual cell changer, a stray light level of less
than 0.01%, and a photometric range of -4 to +4. With a wavelength accuracy of ±0.1 nm
at
656.1 nm, a baseline flatness of ±0.0006 Abs, and a selectable scan speed (slow, medium,
fast), the spectrophotometer ensures precision and flexibility. Additionally, it
supports
variable bandwidth from 0.1 nm to 5 nm, enabling detailed spectral analysis. The
inclusion
of an integrating sphere facilitates the measurement of solid samples, such as metal
oxides
and powders, making it versatile for diverse research and industrial applications.